| Data log functions |
Measure functions |
Voltage, current, resistance, frequency, temperature, pressure |
| Reading rate |
1, 2, 5, 10, 20, 30, or 60 readings/minute |
| Maximum record length |
8000 readings (7980 for 30 or 60 readings/minute) |
| Ramp functions |
Source functions |
Voltage, current, resistance, frequency, temperature |
| Rate |
4 steps/second |
| Trip detect |
Continuity or voltage (continuity detection not available when sourcing current) |
| Loop power function |
Voltage |
Selectable, 26 V |
| Accuracy |
10%, 18 V minimum at 22 mA |
| Maximum current |
25 mA, short circuit protected |
| Maximum input voltage |
50 V DC |
| Step functions |
Source functions |
Voltage, current, resistance, frequency, temperature |
| Manual step |
Selectable step, change with arrow buttons |
| Autostep |
Fully programmable for function, start delay, step value, time per step, repeat |
| Environmental Specifications |
| Operating temperature |
-10°C to +50°C |
| Storage temperature |
-20°C to +60°C |
| Dust/water resistance |
Meets IP52, IEC 529 |
| Operating altitude |
3000 m above mean sea level (9842 ft) |
| Safety Specifications |
| Agency approvals |
CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 |
| Mechanical and General Specifications |
| Size |
136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in) |
| Weight |
1.2 kg (2.7 lb) |
| Batteries |
Internal Battery Pack Li-ion: 7.2V, 4400mAh, 30 Wh |
| Battery life |
>8 hours typical |
| Battery replacement |
Replace without opening calibrator; no tools required |
| Side port connections |
Pressure module connector |
| USB Connector to interface to your PC |
| Digital instrument (HART) connector |
| Connection for optional battery charger/eliminator |
| Data storage capacity |
1 week of calibration procedures results |