| Component |
Specification |
| Accuracy Barometer |
±6 mmHg |
| Accuracy Percent Saturation (Polarographic DO) |
±2 % saturation or ±2% of reading, whichever is greater |
| Accuracy Percent Saturation (RDO) |
±2 % saturation up to 200 % saturation; ±10 % saturation above 200 % saturation |
| Accuracy Polarographic DO |
±0.2 mg/L or ±2% of reading, whichever is greater |
| Accuracy RDO |
±0.1 mg/L up to 8 mg/L; ±0.2 mg/L from 8 to 20 mg/L; ±10% of reading up to 50 mg/L |
| Accuracy Temperature |
±0.1°C |
| Alarm Output |
Set point alarm, high/low limit alarm, calibration due alarm |
| Barometric Pressure Correction |
400 to 850 mmHg, automatic using built-in barometer or manual entry |
| Calibration Modes |
Water-saturated air, air-saturated water, manual (Winkler titration), zero point |
| Dimensions LxWxH |
12 cm x 3 cm x 9.5 cm (4.7'' x 1.2'' x 3.8'') |
| Inputs |
9 pin MiniDIN (RDO optical or polarographic DO sensor with built-in temperature) |
| IP Rating |
IP-54 |
| Measurement Modes |
Dissolved oxygen as % saturation or mg/L with temperature and barometric pressure |
| Range Barometric Pressure |
400 to 850 mmHg |
| Range Percent Saturation (Polarographic DO) |
0.0 to 600.0 % saturation |
| Range Percent Saturation (RDO) |
0.0 to 500.0 % saturation |
| Range Polarographic DO |
0.00 to 90.00 mg/L |
| Range RDO |
0.00 to 50.00 mg/L |
| Range Temperature |
0.0 to 50.0 °C, 32.0 to 122.0 °F |
| Relative Humidity Range |
5 to 85 %, non-condensing |
| Weight |
0.2 kg, 0.5 lbs |